The European Journal of Orthodontics Advance Access originally published online on May 4, 2008
The European Journal of Orthodontics 2008 30(5):483-489; doi:10.1093/ejo/cjm124
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Stability of connected mini-implants and miniplates for skeletal anchorage in orthodontics
Orthodontics, Faculty of Dentistry, The University of Hong Kong, SAR, China
Address for correspondence Professor A. Bakr M. Rabie, Orthodontics, Faculty of Dentistry, The University of Hong Kong, Prince Philip Dental Hospital, 34 Hospital Road, Hong Kong SAR, China, E-mail: rabie{at}hkusua.hku.hk
| Summary |
|---|
|
|
|---|
The aim of this study was to examine the primary stability of connected mini-implants and miniplates. Three different skeletal anchorage systems were investigated: (1) two 1.5 mm diameter cylindrical mini-implants connected with a 0.021 x 0.025 inch stainless steel (SS) wire, (2) two 1.6 mm diameter tapered mini-implants connected with a 0.021 x 0.025 inch SS wire, and (3) two 2.0 mm diameter cylindrical mini-implants connected by a titanium locking miniplate. Fifteen standardized bovine bone specimens were prepared, five specimens for each experimental group. The connected mini-implants were fixed on the bone specimens. The systems underwent uniaxial pull-out tests at the midpoint of the connecting wire or miniplate using a mechanical testing machine. One-way analysis of variance was used to determine the difference of the pull-out test results between the groups.
Both the titanium miniplate and SS wire connection systems showed severe deformation at the screw head, which broke before the mini-implants failed. The 2.0 mm miniplate system showed the highest pull-out force (529 N) compared with the other two wire connection systems (P < 0.001). The 2.0 mm system was also stiffer than the 1.6 and 1.5 mm systems (P < 0.001). The yield force of the 2.0 mm miniplate (153 N) was significantly higher than the 1.5 mm (88 N) and 1.6 mm (76 N) systems (P < 0.001).
This in vitro study demonstrated that the connection of two mini-implants with a miniplate resulted in higher pull-out force, stiffness, and yield force to resist pulling force and deformation. Such a set-up could thus provide a stable system for orthodontic skeletal anchorage.
| Introduction |
|---|
|
|
|---|
Anchorage is fundamental to the success of orthodontic treatment. Use of skeletal anchorage in orthodontics has gained increasing popularity in both clinical applications and research since its introduction (Creekmore and Eklund, 1983
Different types of skeletal anchorage have been adopted for a wide range of applications. The main modalties commonly used are (1) conventional endosseous implants of normal size, (2) mini-implants that are usually less than 2 mm in diameter, and (3) miniplates fixed with miniscrews (Higuchi and Slack, 1991
; Kanomi, 1997
; Sugawara et al., 2002
). The success rate of mini-implants and miniplates is reported to be more than 80 per cent (Miyawaki et al., 2003
; Cheng et al., 2004
; Park et al., 2006
). Primary stability is an important factor in the success of implants. Using two or more connected implants might increase stability (Balshi and Wolfinger, 1997a
, b
) and the connection of two mini-implants has indeed been shown to improve the stability of the skeletal anchorage system (Sugawara et al., 2002
; Chang et al., 2004
; Lee et al., 2004
).
A new miniplate system that has a locking feature could also increase implant stability (Alpert et al., 2003
; Gutwald et al., 2003
). The locking screw has a special double-thread design that allows the threaded screw head to engage the corresponding threaded plate holes on the miniplate during insertion. To date, no published study has sought to biomechanically evaluate the stability of connected mini-implants and miniplates. The objective of this study was to examine the primary stability of connected mini-implants and miniplates.
| Materials and methods |
|---|
|
|
|---|
Bone harvesting and preparation
Fifteen bone plates were harvested from 15 mandibles of recently killed domestic oxen. The soft tissues were stripped off the angle of the mandible. Bone specimens of 30 x 40 mm were cut from one side of the mandible with a bone saw. These were then ground under coolant until 2 mm of cortical bone was left with 6 mm of cancellous bone attached. After preparation, the bone specimens were stored in a solution of equal amounts of normal saline and 70 per cent ethanol at 4°C before mechanical testing was carried out within 1 week (Mittra et al., 2005
).
Pull-out tests were performed on the 15 bone specimens using three types of skeletal anchorage systems (Figure 1): (1) 1.5 x 7 mm (diameter 1.5 mm and length 7 mm) cylindrical mini-implants (Bracket head type, AbsoAnchor, Dentos Inc., Seoul, Korea); (2) 1.6 x 7 mm tapered mini-implants (Regular type, Orlus, Ortholution Co., Seoul, Korea); and (3) 2.0 x 8 mm cylindrical miniplate locking screws (Lock screw; Synthes GmbH, Oberdorf, Switzerland). Each of the three test groups consisted of five bone specimens.
|
Pilot holes were drilled perpendicularly in the bone specimens which were then fixed onto a drilling machine. A 1.2 mm pilot drill with a depth of 4 mm was used for the 1.5 mm mini-implant and the 1.6 mm tapered mini-implant and a 1.5 mm pilot drill with a depth of 5 mm for the 2.0 mm mini-implants. The bone specimens were then secured onto the crosshead platform of a mechanical testing machine (Model 1185; Instron, Norwood, Massachusetts, USA) by a steel plate with a 20 x 30 mm window.
Mechanical testing of connected mini-implants
For the 2.0 mm miniplate system, two mini-implants, 15 mm apart, were connected with a titanium locking plate (Lock Plate, 1 mm thick, straight, four holes, Synthes GmbH) that was offset by 1 mm from the surface of the bone specimen. Threads on both the head of the screw and the inside of the plate holes allowed the locking screws to be secured to the locking plate as they were screwed in (Figure 2). For the 1.5 mm cylindrical mini-implant and 1.6 mm tapered mini-implant, two mini-implants 15 mm apart were inserted into the bone with 1 mm offset after pre-drilling. Subsequently, a section of a 0.021 x 0.025 inch stainless steel (SS) archwire (Kleen PakTM System, Ormco Corporation, Orange, California, USA) was inserted and wrapped around the slot at the screw head with the thicker side in the slot. A 0.010 inch steel ligature wire was tied to fix the archwire at the screw head. The ligature wire, archwire, and screw head were then embedded in composite resin using the TransbondTM XT system (3M Unitek Orthodontic Products, Monrovia, California, USA; Figure 3).
|
|
Uniaxial pull-out force was applied by the testing machine through a rectangular SS blade at the midpoint of the archwire or miniplate (Figures 4 and 5). The crosshead platform speed was 1 mm per minute and testing was terminated when a component in the systems broke and the loading force dropped to zero. From the force–displacement graph plotted by the testing machine, the maximum pull-out force, displacement at the maximum pull-out force, stiffness (gradient of load to displacement), yield load, and yield displacement were evaluated. Yield load and yield displacement were the load and the displacement, respectively, at which permanent deformation began (Haug et al., 2002
|
|
Statistical analysis
One-way analysis of variance was applied to assess the difference of pull-out test results between mini-implants of different groups using the Statistical Package for Social Sciences (Version 13, SPSS Inc., Chicago, Illinois, USA). The level of statistical significance was set at P < 0.05. Multiple comparisons were undertaken using a Bonferroni test at
= 0.05.
| Results |
|---|
|
|
|---|
For the 1.5 mm mini-implant system, which was connected by SS wire, the first response to the pull-out test was a V-shaped deformation of the archwire at the point of force application. Following this, the composite resin and ligature wire fractured on one or both of the mini-implants. The screw heads then broke at the slot in three cases and at the neck in two cases (Figure 6a,b). The remaining parts of the screws remained in the bone and showed only slight deformation or tilting.
|
The response to the pull-out test for the 1.6 mm mini-implant system was initially also deformation at the midpoint of the archwire. Again, the composite resin and ligature wire broke off. The archwire then slipped from the slot of the screw head and the force dropped to zero. The two mini-implants stayed in the bone and were intact with only slight deformation or tilting (Figure 6c).
Deformation with the 2.0 mm miniplate system started at the middle of the miniplate which became more deformed and elongated. Screw threads at the screw head and the inside of the holes of the miniplate stripped off. As a result, the miniplate slipped off the screws and testing was terminated (Figure 6d). The damaged mini-implants were left in the bone and showed more distortion and tilting than the other two systems (Figure 6e).
The force–displacement graphs (Figure 7a–d) show the loading behaviour of the three systems. With the 2.0 mm system, the miniplate kept deforming and elongating until the miniplate broke off from one of the mini-implants and then the force dropped to zero. In the other two wire connection systems, there were two to three sudden decreases in force level before a decrease to zero. Breakage of the resin or wire on one or both of the mini-implants corresponded to the sudden decrease on the force–displacement graph.
|
Comparison of the pull-out test results of the three systems (Table 1) showed that the maximum pull-out force of the titanium miniplate in the 2.0 mm mini-implants was 529 N which was significantly higher than the 1.6 mm (374 N) and 1.5 mm (316 N) mini-implants connected with SS wire (P < 0.001). There were significant differences in displacement between the 2.0 and 1.6 mm groups and between the 2.0 and 1.5 mm groups (P < 0.001). The titanium miniplate in the 2.0 mm mini-implant system was stiffer than the 1.6 and 1.5 mm systems (P < 0.001). No difference was found between the 1.6 and 1.5 mm groups (P > 0.05). The yield force of the 2.0 mm miniplate system (153 N) was significantly higher than that of the 1.5 mm (88 N) and 1.6 mm (76 N) systems (P < 0.001). The yield displacement was significantly larger in the 1.5 mm system than in the 2.0 mm miniplate system (P < 0.001).
|
| Discussion |
|---|
|
|
|---|
This study showed that the connection of two mini-implants with a miniplate provided a stable structure to resist pulling force and deformation. There are case reports in the literature investigating the effects of connecting two mini-implants for increased stability for skeletal anchorage (Chang et al., 2004
In the 1.6 mm system, the force was transmitted to the connecting parts, which were the composite resin and ligature wire. These broke before the force was transmitted to the mini-implants, hence less tilting was seen. In the 1.5 mm system, the mini-implant itself broke at the slot or the neck as well as the connecting parts. This could be because of the smaller dimensions of the components at the slot and the neck of the mini-implants in this system compared with the other two systems.
The miniplate system also had the highest stiffness, 266 N/mm, compared with the 1.6 and 1.5 mm systems, which had a stiffness of 128 and 98 N/mm, respectively. The stiffness is the force needed to deform the system per millimetre. This stiffness should be interpreted in conjunction with the yield load and yield displacement. The yield load of the miniplate system was 153 N and the yield displacement 0.68 mm. This means that permanent deformation of the miniplate system began only when the force was as high as 153 N and when the system was deformed more than 0.68 mm. To deform the miniplate system by 1 mm, 266 N of force was needed. The 1.5 and 1.6 mm systems deformed when the force was 88 and 76 N, respectively, as indicated by their yield forces. Normally, the magnitude of orthodontic force is less than 5 N but the average human bite force could be as high as 200 N. The average force transmitted to a bracket during mastication has been reported to be between 40 and 120 N (Powers et al., 1997
; Paphangkorakit and Osborn, 1998
). The findings give some insight into the possible force range that a mini-implant might need to withstand in the oral environment. The yield load of 153 N of the miniplate system suggests that it would withstand this range of intraoral masticatory force. The force applied with a rapid maxillary expander can be up to 120 N (Sander et al., 2006
). The 153 N yield load of the miniplate may thus also withstand this range of force.
In this experiment, the loading force for the mechanical testing was applied in a flatwise manner, i.e. the force was perpendicular to the flat surface. The force needed to deform the system in a flatwise direction is much lower than that required in an edgewise manner (Hegtvedt et al., 1994
; Loukota and Shelton, 1995
). Thus, the yield force to permanently deform the miniplate would be higher if the miniplate was not loaded flatwise but in other directions. In clinical application, the skeletal anchorage system is usually not loaded in a flatwise direction and a higher yield force would be expected.
The connection of two mini-implants may not only provide a stable anchorage system but also improve the versatility of the device as well. The relationship between the occlusogingival position of the mini-implants and centre of resistance of the teeth to be moved is important in controlling the mode of movement of teeth (Park et al., 2005
). Application of force from different positions along the miniplate connector may allow the direction or mode of tooth movement during treatment to be changed from tipping to torquing, or vice versa.
The screw threads on the inside of the holes of the miniplate may allow different customized devices to be screwed onto it. As a result, different attachments could be screwed onto the miniplate, as an alternative to replacement of individual mini-implants with different screw heads. Even an orthodontic attachment such as a bracket may be directly screwed onto the system which could have individualized torque and rotation control without unwanted movement of other teeth (Kyung et al., 2005
).
The findings showed that the strength and design of the connecting part were vital to the anchorage device. If the connecting plate and the screw head were made of stronger materials, e.g. SS or a stronger metal alloy, the stability of the anchorage device might be further increased. Having a double-thread design at the screw head is one design which is advantageous in connecting mini-implants but this is not the only one. Further research needs to be conducted to develop more convenient and stronger screw head designs to connect mini-implants for improved stability.
| Conclusion |
|---|
|
|
|---|
This is the first in vitro study to demonstrate that the connection of two mini-implants with a miniplate results in higher pull-out force, stiffness, and yield force to resist pulling force and deformation. Such a set-up could thus provide a stable system for orthodontic skeletal anchorage.
| Funding |
|---|
|
|
|---|
The University Strategic Research Theme: Genomics, Proteomics and Bioinformatics, University of Hong Kong (10206152.11222.21700.302 [EC] .01); a Competative Earmarked Research Grant (CERG) grant (102006968.22311.08003.324 [EC] .01).
| Acknowledgement |
|---|
We would like to thank Dr Trevor Lane, Dr Henry Luk, Mr Tony Yuen, Mr C. Y. Yip, and Mr Shadow Yeung for their suggestions and technical support in the research.
| References |
|---|
|
|
|---|
-
Alpert B, Gutwald R, Schmelzeisen R. New innovations in craniomaxillofacial fixation: the 2.0 lock system. Keio Journal of Medicine (2003) 52:120–127.[Medline]
Balshi TJ, Wolfinger GJ. Immediate loading of Brånemark implants in edentulous mandibles: a preliminary report. Implant Dentistry (1997a) 6:83–88.[Medline]
Balshi TJ, Wolfinger GJ. Two-implant-supported single molar replacement: interdental space requirements and comparison to alternative options. International Journal of Periodontics and Restorative Dentistry (1997b) 17:426–435.
Chang YJ, Lee HS, Chun YS. Microscrew anchorage for molar intrusion. Journal of Clinical Orthodontics (2004) 38:325–330.[Medline]
Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. International Journal of Oral and Maxillofacial Implants (2004) 19:100–106.
Creekmore TD, Eklund MK. The possibility of skeletal anchorage. Journal of Clinical Orthodontics (1983) 17:266–269.[Medline]
Gutwald R, Alpert B, Schmelzeisen R. Principle and stability of locking plates. Keio Journal of Medicine (2003) 52:21–24.[Medline]
Haug RH, Peterson GP, Goltz M. A biomechanical evaluation of mandibular condyle fracture plating techniques. Journal of Oral and Maxillofacial Surgery (2002) 60:73–80.[CrossRef][Web of Science][Medline]
Hegtvedt AK, Michaels GC, Beals DW. Comparison of the resistance of miniplates and microplates to various in vitro forces. Journal of Oral and Maxillofacial Surgery (1994) 52:251–257.[Web of Science][Medline]
Henry PJ, Singer S. Implant anchorage for the occlusal management of developmental defects in children: a preliminary report. Practical Periodontics and Aesthetic Dentistry (1999) 11:699–706.[Medline]
Higuchi KW, Slack JM. The use of titanium fixtures for intraoral anchorage to facilitate orthodontic tooth movement. International Journal of Oral and Maxillofacial Implants (1991) 6:338–344.
Kanomi R. Mini-implant for orthodontic anchorage. Journal of Clinical Orthodontics (1997) 31:763–767.[Medline]
Kircelli BH, Pektas ZO, Uckan S. Orthopedic protraction with skeletal anchorage in a patient with maxillary hypoplasia and hypodontia. Angle Orthodontist (2006) 76:156–163.[Web of Science][Medline]
Kyung SH, Choi HW, Kim KH, Park YC. Bonding orthodontic attachments to miniscrew heads. Journal of Clinical Orthodontics (2005) 39:348–353.[Medline]
Lee JS, Kim DH, Park YC, Kyung SH, Kim TK. The efficient use of midpalatal miniscrew implants. Angle Orthodontist (2004) 74:711–714.[Web of Science][Medline]
Loukota RA, Shelton JC. Mechanical analysis of maxillofacial miniplates. British Journal of Oral and Maxillofacial Surgery (1995) 33:174–179.[CrossRef]
Mittra E, Rubin C, Qin YX. Interrelationship of trabecular mechanical and microstructural properties in sheep trabecular bone. Journal of Biomechanics (2005) 38:1229–1237.[CrossRef][Web of Science][Medline]
Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. American Journal of Orthodontics and Dentofacial Orthopedics (2003) 124:373–378.[CrossRef][Web of Science][Medline]
Paphangkorakit J, Osborn JW. Effects on human maximum bite force of biting on a softer or harder object. Archives of Oral Biology (1998) 43:833–839.[CrossRef][Web of Science][Medline]
Park HS, Bae SM, Kyung HM, Sung JH. Simultaneous incisor retraction and distal molar movement with microimplant anchorage. World Journal of Orthodontics (2004a) 5:164–171.[Medline]
Park HS, Kwon TG, Sung JH. Nonextraction treatment with microscrew implants. Angle Orthodontist (2004b) 74:539–549.[Web of Science][Medline]
Park HS, Kwon OW, Sung JH. Microscrew implant anchorage sliding mechanics. World Journal of Orthodontics (2005) 6:265–274.[Medline]
Park HS, Jeong SH, Kwon OW. Factors affecting the clinical success of screw implants used as orthodontic anchorage. American Journal of Orthodontics and Dentofacial Orthopedics (2006) 130:18–25.[CrossRef][Web of Science][Medline]
Powers JM, Kim HB, Turner DS. Orthodontic adhesives and bond strength testing. Seminars in Orthodontics (1997) 3:147–156.[CrossRef][Medline]
Sander C, Huffmeier S, Sander FM, Sander FG. Initial results regarding force exertion during rapid maxillary expansion in children. Journal of Orofacial Orthopedics (2006) 67:19–26.[CrossRef][Medline]
Sugawara J, et al. Treatment and posttreatment dentoalveolar changes following intrusion of mandibular molars with application of a skeletal anchorage system (SAS) for open bite correction. International Journal of Adult Orthodontics and Orthognathic Surgery (2002) 17:243–253.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||






