An in vitro investigation of the influence of self-ligating brackets, low friction ligatures, and archwire on frictional resistance
Department of Oral Science, University G DAnnunzio, Chieti, Pescara, Italy
Address for correspondence Dr Simona Tecco, Department of Oral Science, University G DAnnunzio, Via Le Mainarde 26, Chieti, Pescara 65121, Italy E-mail: simtecc{at}tin.it
| Summary |
|---|
|
|
|---|
This study, performed using a specially designed apparatus that included 10 aligned brackets, evaluated the frictional resistance generated by conventional stainless steel (SS) brackets (Victory Series), self-ligating Damon SL II brackets, Time Plus brackets, and low-friction ligatures (Slide) coupled with various SS, nickel-titanium (NiTi), and beta-titanium (TMA) archwires. All brackets had a 0.022-inch slot and the orthodontic wire alloys were 0.016, 0.016 x 0.022, and 0.019 x 0.025 inch NiTi, 0.017 x 0.025 inch TMA, and 0.019 x 0.025 inch SS. Each bracket–archwire combination was tested 10 times.
Coupled with 0.016 inch NiTi, Victory brackets generated the most friction and Damon SL II the least (P < 0.001); with 0.016 x 0.022 inch NiTi, the self-ligating brackets (Time and Damon SL II) generated significantly lower friction (P < 0.001) than Victory Series and Slide ligatures; with 0.019 x 0.025 inch SS or 0.019 x 0.025 inch NiTi, Slide ligatures generated significantly lower friction than all other groups. No difference was observed among the four groups when used with a 0.017 x 0.025-inch TMA archwire.
These findings suggest that the use of an in vitro testing model that includes 10 brackets provides information about the frictional force of the various bracket–archwire combinations.
| Introduction |
|---|
|
|
|---|
Friction is the resistance to motion when an object moves tangentially against another (Besancon, 1985
However, low friction may be desired during the early alignment phase, when all the teeth move at the same time and the wire slides through 10 brackets and two tubes. For this, a new experimental method to investigate the friction generated during the sliding of an archwire along a group of 10 aligned brackets was introduced (Tecco et al., 2005
).
Self-ligating brackets (Stolzenberg, 1935
, 1946
) are ligature-less bracket systems that have a mechanical device built into the bracket to close off the edgewise slot (Cacciafesta et al., 2003
); they are generally more comfortable for the patients, because of the absence of a wire ligature (Shivapuja and Berger, 1994
) and because they also do not require significant chair side time (Maijer and Smith, 1990
; Damon, 1998
). Several studies have demonstrated a significant decrease in friction using self-ligating brackets compared with a conventional bracket design (Berger, 1990
; Sims et al., 1993
, 1994
; Read-Ward et al., 1997
; Thomas et al., 1998
; Thorstenson and Kusy, 2001
; Khambay et al., 2004
; Henao and Kusy, 2005
).
Recently, as it has been stated that friction is determined mostly by the nature of the ligation (Cacciafesta et al., 2003
), new low-friction ligatures (Slide©, Leone, Firenze, Italy) have been introduced, similar to elastic ligatures, but with an anterior part that is more rigid and similar to the mechanical device of self-ligating brackets. This is useful when low friction is desired, while common ligatures (not low friction) can be used when more friction is required (Figure 1). According to the manufacturer, Slide© is constructed from a special polyurethane mix for medical use. It can be applied in the same way as classical elastic ligatures and, once on the bracket, it self ligates on the slot leaving the wire free to slide and to act on the dentoalveolar structures. According to the manufacturer, its particular form noticeably improves patient comfort during the first phases of treatment.
|
The purpose of this study was to compare the frictional forces generated by two types of self-ligating brackets and conventional stainless steel (SS) brackets, and elastic or low-friction ligatures, using a specially custom-designed apparatus that included 10 brackets.
| Material and methods |
|---|
|
|
|---|
Mechanical testing
The brackets tested were: (1) Damon SL II brackets (SDS, Ormco, Glendora, California, USA), (2) Time Plus brackets (American Orthodontics, Sheboygan, Wisconsin, USA), (3) Victory brackets (Victory Series, 3M Unitek, Monrovia, California, USA) ligated with elastic modules (Ligature Ringlet, RMO, Denver, Colorado, USA), and (4) Victory brackets (3M Unitek) ligated with Slide© low-friction ligatures.
The testing model (Figure 2; Myrmex Laboratory, Foggia, Italy) has been described previously (Tecco et al., 2005
).
|
On the test model, 10 brackets of the same group were mounted in alignment using a cyanoacrylate adhesive (Loctite 416, Loctite Corp., Rocky Hill, Connecticut, USA). The alignment of the brackets was obtained through preliminary insertion of a 0.021 x 0.028-inch SS archwire in the slots of the brackets, without ligation, before bonding. After bonding of the brackets on the metal bar, the SS archwire was removed. However, as minor malalignments of the brackets or non-linearity of the wire could not be controlled, to estimate the extent to which the friction could be attributed to malalignment rather than ligation, a confirmatory check was performed by measuring the friction for each bracket–archwire combination with only the terminal brackets ligated.
In total, 20 test models were constructed, i.e. five models for each group of brackets (Victory Series, Time, Damon SL II and Slide ligatures). For each group, a single model was used 10 times to test the same bracket–archwire combination with all the brackets ligated, and 10 times to test the same bracket–archwire combination with only the terminal brackets ligated. In this way, each model was used for a total of 20 tests.
The archwires (Table 1) were selected for testing as a representation of those used in various stages of orthodontic treatment: 0.016, 0.016 x 0.022, and 0.019 x 0.025 inch nickel-titanium (NiTi), 0.017 x 0.025 inch beta-titanium (TMA), and 0.019 x 0.025 inch SS. For each testing procedure, a new archwire was employed. The sliding force values (F) were evaluated for each archwire 10 times. Thus, a total of 400 testing procedures were performed. The tests were run in the dry state at an ambient temperature of 34°C.
|
For friction evaluation, a mechanical testing machine (Model 30K, Lloyd Instruments Ltd, East Fareham, Harts, UK) with a 10-lb tension load cell, set on a range of 1 lb and calibrated from 0 to 1000 g, was employed (Figure 2). The archwires were gripped by crimping brass fittings on the distal ends, which allowed sliding of the wire along the 10 brackets and recording of the frictional forces.
The archwires moved through all 10 brackets at a crosshead speed of 0.5 mm/minute. Once archwire movement began, each run lasted approximately 5 minutes. F value were calculated in centi-Newtons (cN). After each test, the machine was stopped, the bracket and wire assembly was removed, and a new assembly was placed. This was undertaken for 20 non-repeated evaluations (10 evaluations with all the brackets ligated and 10 evaluations with only the terminal brackets ligated) for each bracket–archwire combination. A randomized sequence for each type of archwire was performed.
The load cell registered the force levels needed to move the wire along the 10 aligned brackets, and the levels were transmitted to a computer.
Statistical analysis
Descriptive statistics, including the mean and standard deviation (SD), were calculated for each bracket–archwire combination.
In order to estimate the extent to which friction could be attributed to the malalignment of the brackets rather than the type of ligation, for each bracket–archwire combination, a statistical comparison, undertaken using a Student's t-test, was performed between the data for all ligated brackets and the data with only the terminal brackets ligated. In addition, in order to evaluate whether the differences in the friction could be attributed to the malalignment of the brackets, rather than ligation, a one-way analysis of variance (ANOVA) test was carried out to determine the existence of statistically significant differences among the four groups of brackets for each archwire tested, when only the terminal brackets were ligated.
The data were then analysed as the differences in frictional force observed in the five groups of archwires among the four groups of brackets. If the one-way ANOVA test was significant, a t-test for independent groups was calculated to evaluate the significance of the differences among the groups.
To better understand the influence associated with the type of archwire alloy, the data of each type of archwires were then considered (as a mean of the values from the four groups of brackets) and statistically compared using the one-way ANOVA and, if significant, the t-test for between-group comparisons. Values that were not significant were defined as P > 0.05.
| Results |
|---|
|
|
|---|
For each bracket–archwire combination, the F values obtained when all 10 brackets had been ligated were statistically significantly higher than when only the terminal brackets had been ligated (P < 0.001; Table 2). Only for one combination, was no significant difference observed, i.e. when the Damon SL II brackets were engaged with the 0.016-inch NiTi archwire (Table 2).
|
In addition, when the difference among the four types of ligation for each archwire, wase evaluated, no statistically significant difference was observed among the four groups for each archwire (Table 2).
The frictional forces (F), observed when all the brackets had been ligated, for each bracket–archwire combination are reported in Table 2 as the mean and SD, and their significant differences are shown in Figure 3.
|
When engaged with 0.016-inch NiTi wire, the Damon SL II brackets showed significantly lower friction compared with all other groups, while Victory Series brackets showed significantly higher friction. No significant difference was observed between Time brackets and Slide ligatures (Table 2, Figure 3). When engaged with 0.016 x 0.022-inch NiTi wire, the self-ligating brackets (Time and Damon SL II) generated significantly lower friction than Victory Series brackets and Slide ligatures (Table 2, Figure 3). For 0.017 x 0.025 inch TMA, no significant difference was observed among the four groups (Table 2 and Figure 3). With 0.019 x 0.025 inch NiTi and SS, the trend was the same, as Slide ligatures generated significantly lower friction, compared with the other groups, with no significant differences between groups (Table 2, Figure 3).
When comparisons among the different types of archwires were performed, the thicker rectangular archwires (0.017 x 0.025 inch TMA, 0.019 x 0.025 inch SS, and 0.019 x 0.025 inch NiTi) showed a significantly higher level of friction when compared with 0.016 and 0.016 x 0.022 inch NiTi. No other significant difference was observed (Figure 4).
|
| Discussion |
|---|
|
|
|---|
Force magnitude during orthodontic treatment will result in optimal tissue response and rapid tooth movement (Cacciafesta et al., 2003
The model employed in this investigation
In this investigation, the use of a model that includes 10 aligned brackets provides more data than previous studies (Andreasen and Quevedo, 1970
; Kapila et al., 1990
; Rose and Zernik, 1996
; Braun et al., 1999
; Cacciafesta et al., 2003
; Khambay et al., 2004
, 2005
) or a typodont (Henao and Kusy, 2005
) and provides interesting observations. Some recent studies have used different methods to test the frictional force between bracket and archwires; these other methods are briefly described, since the differences in methodology should be acknowledged before analysis and comparison of the results.
Henao and Kusy (2005)
, for example, used typodont models either mounted with a self-ligating or a conventional design. The authors observed that, at the third-stage archwires, no archwire was engaged, probably because of the minor malalignments of the brackets on the typodont. Thus, for the present investigation, a linear design of the brackets was chosen.
In another recent study, a model which employed only one Damon SL II bracket and one conventional straightwire bracket was used (Khambay et al., 2004
). In that investigation, 0.017 x 0.025- and 0.019 x 0.025-inch SS and TMA wires were tested with four types of elastomeric modules and preformed 0.09-inch SS ligatures. Each bracket/wire combination with each method of ligation was tested 10 times in the presence of human saliva. The authors found that the Damon II self-ligating and unligated conventional SS brackets produced negligible mean frictional forces with any of the wires tested, but for the 0.017 x 0.025- and 0.019 x 0.025-inch SS or 0.019 x 0.025-inch TMA wires, SS ligatures produced the lowest mean frictional forces.
Khambay et al. (2005)
used a maxillary premolar bracket (3M Unitek) that was welded to a sheet of SS and glued to a Perspex block. The bracket base was removed and the cut continued into the Perspex below. A length of test wire was taken and bent to form a U shape, with the middle portion 20 mm in length. The free ends of the wire were secured to the load cell of the Nene testing machine. Two wire sizes were tested, 0.017 x 0.025 and 0.019 x 0.025 inch SS. The load cell was activated and the force with which the wire was displaced into the slot by the ligation method was measured. However, while the complex assembly of brackets used in that investigation seems to offer much greater potential, the results could be affected by minor malalignments of the brackets or non-linearity of the wire, thereby invalidating any conclusions on the effects of ligation.
For these reasons, in the present study, the brackets were accurately aligned during assembly using the 0.022 x 0.028-inch SS wire. In addition, in order to provide an estimate of the extent to which the friction could be attributed to malalignment rather than ligation, frictional forces were also recorded when only the terminal brackets had been ligated and the data were compared with frictional forces observed when all 10 brackets had been ligated, for each bracket–archwire combination.
Finally, another confirmatory check was performed as the existence of differences among the four groups of ligations was evaluated for each archwire employed, when only the terminal brackets had been ligated.
As noted, for each bracket–archwire combination, the friction values obtained when all 10 brackets had been ligated resulted in statistically significantly higher friction than when only the terminal brackets had been ligated (P < 0.001; Table 2). Only for the Damon SL II brackets engaged with 0.016-inch NiTi archwire was no significant difference observed (Table 2). These observations seem to suggest that ligation plays an important role in the generation of friction between bracket and slot. The result for Damon SL II, when engaged with 0.016-inch NiTi archwire is in agreement with the findings of Khambay et al. (2004)
, with a model that included only one bracket.
When the differences among the four groups, with only the terminal brackets ligated, were evaluated, no statistically significant difference was found among the groups for each archwire (Table 2), suggesting that all the differences observed when all the 10 brackets had been ligated (Figures 3 and 4) could be attributed to the type of ligation.
With regard to the testing model employed in this study, it must be noted that the archwires moved through all 10 brackets at a crosshead speed of 0.5 mm/minute and once archwire movement began, each run lasted approximately 5 minutes. This crosshead speed is generally lower than that used in other investigations. In the study by Khambay et al. (2004)
, which included one bracket, the specimens were tested on a Nene M3000 testing machine, with a crosshead speed of 5 mm/minute and each test run lasted 4 minutes. The speed used in the present investigation was generally lower than that in other studies, since the 10-bracket model needed very small movements of the archwires in order to avoid failure of the brackets from the metal bar.
Findings
One of the main findings of the present study was that self-ligating and Victory Series brackets showed a different trend depending on the section (round or rectangular) of the archwire (Figure 3). While Victory brackets generated significantly higher friction when coupled with 0.016 inch NiTi compared with the two self-ligating brackets (Damon SL II and Time), no significant differences among Victory Series, Damon SL II, and Time brackets were observed when engaged with the rectangular archwires (Figure 3). This seems to indicate that the design of the self-ligating brackets results in low friction only when engaged with round wires, and not with rectangular archwires.
However, it should be noted that Damon SL II showed the lowest level of friction with round wires, compared with all the other combinations, suggesting that, among the considered archwire–bracket combinations it remains the bracket of choice when lower frictional force is required during the alignment phase.
Another interesting observation concerns the results observed with Slide ligatures (Figure 3). They showed similar friction compared with self-ligating brackets when coupled with 0.016 inch NiTi, 0.016 x 0.022 inch NiTi, and 0.017 x 0.025 inch TMA, that is low friction with round wires and high friction with rectangular wires. In addition, when coupled with 0.016 x 0.022 inch NiTi, they also generated significantly higher friction compared with the two self-ligating brackets. However, when engaged with 0.019 x 0.025-inch NiTi and SS archwires, their trend changed as they generated significantly lower friction compared with all other groups. It was hypothesized that this could be associated with their elastic design; probably, when coupled with the larger archwires, their elastic properties decrease and they may loose the capability to create high friction between the archwire and slot. The difference observed could probably be explained by the design of self-ligating brackets. Self-ligating bracket systems have a mechanical device built into the bracket to close off the edgewise slot; this device generally has a small metal sliding door that is rigid and rather stiff, compared with the soft and elastic surface of the low-friction ligatures.
Clinically, the lower friction observed with the Slide ligatures when coupled with rectangular wires could be considered either as an advantage or as a disadvantage in different situations; for example, during anterior tooth retraction, lower friction is required in the lateral segment of the dental arches, while in the final phase of stabilization, higher friction is desired in all slots. The primary advantage of Slide ligatures seems to be that they can be used when low friction is necessary. These findings seem to confirm one of the manufacture's statements, i.e. that Slide ligatures must be used only when low friction is needed.
Another important finding of the present study was that there was no significant difference in frictional force between the two groups of self-ligating brackets; only a significant difference was observed when coupled with 0.016 inch NiTi. This seems to suggest that, despite their differences in structural design and material composition of the bracket slot and cap (Shivapuja and Berger, 1994
), the level of the generated frictional force is approximately the same, except for the significantly lower friction generated by Damon SL II when coupled with 0.016 inch NiTi.
The findings of the present investigation are not in complete agreement with the general statement that SS self-ligating brackets generate lower frictional resistance than conventional SS brackets (Sims et al., 1993
, 1994
; Shivapuja and Berger, 1994
; Pizzoni et al., 1998
; Thomas et al., 1998
; Thorstenson and Kusy, 2001
; Cacciafesta et al., 2003
; Tecco et al., 2005
) as the friction level changed when different archwire sizes and shapes were employed, but no difference in friction was observed among the self-ligating and conventional SS brackets, when coupled with 0.017 x 0.025 inch TMA, 0.019 x 0.025 inch NiTi, and 0.019 x 0.025 inch SS. Thus, the self-ligating brackets did not appear to have any advantages when larger rectangular archwires were used. This observation seems to indirectly indicate that previous studies had often failed to show dramatic reductions in treatment time when self-ligating brackets were used (Harradine, 2001
; Cacciafesta et al., 2003
).
The present study also demonstrated that the wire alloy and the size and shape of its section seem to have a significant influence on friction, as 0.017 x 0.025 inch TMA, 0.019 x 0.025 inch NiTi, and 0.019 x 0.025 inch SS showed a significantly higher frictional force when compared with 0.016 and 0.016 x 0.022-inch NiTi archwires, suggesting that, generally, larger rectangular archwires generate higher friction than round small archwires.
However, although this concept is generally clinically accepted, there were some differences between the present findings and those of other researchers; for example, in this study no significant difference was observed among TMA, SS, and NiTi archwires in terms of friction, although it has been reported that TMA generates higher friction than both SS and NiTi for all bracket–archwire combinations (Angolkar et al., 1990
; Pratten et al., 1990
; Ireland et al., 1991
; Keith et al., 1993
; Downing et al., 1994
; Dickson and Jones, 1996
; Bazakidou et al., 1997
; Loftus et al., 1999
; Cacciafesta et al., 2003
) probably due to the adherence of the archwire material to the slot material during the experimental procedure (Kusy and Whitley, 1990
; Cacciafesta et al., 2003
).
However, in agreement with the findings of Loftus et al. (1999)
and Cacciafesta et al. (2003)
, but not with Angolkar et al. (1990)
who found higher frictional forces with SS, no significant differences were observed between NiTi and SS archwires in terms of frictional force.
This general variability in the reported findings may be due to differences in the experimental set-up or the number of brackets or angulations between the bracket and wire, which in many studies was not zero (Ogata et al., 1996
). Therefore, direct comparison of the various published studies on this topic is complex.
One limitation of the present research, is that it was carried out under ideal conditions, in a passive, not an active configuration which included some malalignment of brackets, as shown in previous reports (Angolkar et al., 1990
; Bednar et al., 1991
; Keith et al., 1993
; Downing et al., 1994
; Shivapuja and Berger, 1994
; Bazakidou et al., 1997
; Thomas et al., 1998
; Loftus et al., 1999
; Cacciafesta et al., 2003
; Tecco et al., 2005
). Frictional investigations in an active configuration (with different bracket angulations) are still in progress and it will be useful in the future to compare these findings with those in the passive state.
Another limitation of this in vitro testing model concerns the lack of reproducibility of tipping, that always occurs when orthodontic force is applied to a tooth, even when a fixed appliance is used.
In addition, the functional forces of the stomatognathic muscles, which could affect tooth orthodontic movement and the effect of the saliva were not considered. These limitations could have influenced the findings.
| Conclusion |
|---|
|
|
|---|
When coupled with 0.016-inch NiTi wire, the Damon SL II brackets showed significantly lower friction compared with all other groups, while Victory Series brackets showed significantly higher friction. With 0.016 x 0.022 inch NiTi, the self-ligating brackets (Time and Damon SL II) generated significantly lower friction than Victory Series brackets and Slide ligatures, while, with 0.019 x 0.025 inch NiTi and 0.019 x 0.025 inch SS, Slide ligatures generated significantly lower friction, compared with the other groups. There was, however, no significant difference among the other groups.
When comparisons among the different types of archwires were performed, the thicker rectangular archwires (0.017 x 0.025 inch TMA, 0.019 x 0.025 inch SS, 0.019 x 0.025 inch NiTi) showed a significantly higher level of frictional force when compared with 0.016 inch and 0.016 x 0.022 inch NiTi.
| References |
|---|
|
|
|---|
-
Andreasen GF, Quevedo FR. Evaluation of frictional forces in the 0.022 x 0.028 edgewise bracket in vitro. Journal of Biomechanics (1970) 3:151–160.[CrossRef][Web of Science][Medline]
Angolkar PV, Kapila S, Duncanson MG Jr, Nanda RS. Evaluation of friction between ceramic brackets and orthodontic wires of four alloys. American Journal of Orthodontics and Dentofacial Orthopedics (1990) 98:499–506.[Web of Science][Medline]
Bazakidou E, Nanda RS, Duncanson MG Jr, Sinha P. Evaluation of frictional resistance in esthetic brackets. American Journal of Orthodontics and Dentofacial Orthopedics (1997) 112:138–144.[CrossRef][Web of Science][Medline]
Bednar JR, Gruendeman GW, Sandrik JL. A comparative study of frictional forces between orthodontic brackets and arch wires. American Journal of Orthodontics and Dentofacial Orthopedics (1991) 100:513–522.[Web of Science][Medline]
Berger JL. The influence of the Speed bracket's self-ligating design on force levels in tooth movement: a comparative in vitro study. American Journal of Orthodontics and Dentofacial Orthopedics (1990) 97:219–228.[Web of Science][Medline]
Besancon RM. The encyclopedia of physics (1985) 3rd edn. New York: Van Nostrand Reinhold Company.
Braun S, Bluestein M, Moore BK, Benson G. Friction in perspective. American Journal of Orthodontics and Dentofacial Orthopedics (1999) 115:619–627.[CrossRef][Web of Science][Medline]
Cacciafesta V, Sfondrini MF, Ricciardi A, Scribante A, Klersy C, Auricchio F. Evaluation of friction of stainless steel and esthetic self-ligating brackets in various bracket-archwire combinations. American Journal of Orthodontics and Dentofacial Orthopedics (2003) 124:395–402.[CrossRef][Web of Science][Medline]
Damon DH. The Damon low-friction bracket: a biologically compatible straight-wire system. Journal of Clinical Orthodontics (1998) 32:670–680.[Medline]
Dickson J, Jones S. Frictional characteristics of a modified ceramic bracket. Journal of Clinical Orthodontics (1996) 30:516–518.[Medline]
Downing A, McCabe J, Gordon P. A study of frictional forces between orthodontic brackets and archwires. British Journal of Orthodontics (1994) 21:349–357.[Abstract]
Harradine NW. Self-ligating brackets and treatment eficiency. Clinical Orthodontics and Research (2001) 4:220–227.[CrossRef][Medline]
Henao SP, Kusy RP. Frictional evaluations of dental typodont models using four self-ligating designs and a conventional design. The Angle Orthodontist (2005) 75:75–85.[Web of Science][Medline]
Ireland AJ, Sherriff M, McDonald F. Effect of bracket and wire composition on frictional forces. European Journal of Orthodontics (1991) 13:322–328.
Kapila S, Angolkar PV, Duncanson MG Jr, Nanda RS. Evaluation of friction between edgewise stainless steel brackets and orthodontic wires of four alloys. American Journal of Orthodontics and Dentofacial Orthopedics (1990) 98:117–126.[Web of Science][Medline]
Keith O, Jones SP, Davies EH. The influence of bracket material, ligation force and wear on frictional resistance of orthodontic brackets. British Journal of Orthodontics (1993) 20:109–115.[Abstract]
Khambay B, Millett D, McHugh S. Evaluation of methods of archwire ligation on frictional resistance. European Journal of Orthodontics (2004) 26:327–332.
Khambay B, Millett D, McHugh S. Archwire seating forces produced by different ligation methods and their effect on frictional resistance. European Journal of Orthodontics (2005) 27:302–308.
Kusy RP, Whitley JQ. Coefficients of friction for arch wires in stainless steel and polycrystalline alumina bracket slots: I. The dry state. American Journal of Orthodontics and Dentofacial Orthopedics (1990) 98:300–312.[Web of Science][Medline]
Loftus BP, Årtun J, Nicholls JI, Alonzo TA, Stoner JA. Evaluation of friction during sliding tooth movement in various bracket-archwire combinations. American Journal of Orthodontics and Dentofacial Orthopedics (1999) 116:336–345.[CrossRef][Web of Science][Medline]
Maijer R, Smith DC. Time savings with self-ligating brackets. Journal of Clinical Orthodontics (1990) 24:29–31.[Medline]
Ogata RH, Nanda RS, Duncanson MG Jr, Sinha PK, Currier GF. Frictional resistances in stainless steel bracket-wire combinations with effects of vertical deflections. American Journal of Orthodontics and Dentofacial Orthopedics (1996) 109:535–542.[CrossRef][Web of Science][Medline]
Pizzoni L, Ravnholt G, Melsen B. Frictional forces related to self-ligating brackets. European Journal of Orthodontics (1998) 20:283–291.
Pratten DH, Popli K, Germane N, Gunsolley JC. Frictional resistance of ceramic and stainless steel orthodontic brackets. American Journal of Orthodontics and Dentofacial Orthopedics (1990) 98:398–403.[CrossRef][Web of Science][Medline]
Read-Ward GE, Jones SP, Davies EH. A comparison of self-ligating and conventional orthodontic bracket systems. British Journal of Orthodontics (1997) 24:309–317.[Abstract]
Rose CM, Zernik JH. Reduced resistance to sliding in ceramic brackets. Journal of Clinical Orthodontics (1996) 30:78–84.[Medline]
Shivapuja PK, Berger J. A comparative study of conventional ligation and self-ligation bracket system. American Journal of Orthodontics and Dentofacial Orthopedics (1994) 106:472–480.[Web of Science][Medline]
Sims APT, Waters NE, Birnie DJ, Pethybridge RJ. A comparison of the forces required to produce tooth movement in vitro using two self-ligating brackets and a pre-adjusted bracket employing two types of ligation. European Journal of Orthodontics (1993) 15:377–385.
Sims APT, Waters NE, Birnie DJ. A comparison of the forces required to produce tooth movement ex vivo through three types of pre-adjusted brackets when subjected to determined tip or torque values. British Journal of Orthodontics (1994) 21:367–373.[Abstract]
Stolzenberg J. The Russell attachment and its improved advantages. International Journal of Orthodontia and Dentistry for Children (1935) 21:837–840.[CrossRef]
Stolzenberg J. The efficiency of the Russell attachment. American Journal of Orthodontics and Oral Surgery (1946) 32:572–582.[CrossRef]
Tecco S, Festa F, Caputi S, Traini T, Di Iorio D, D'Attilio M. Friction of conventional and self-ligating brackets using a 10 bracket model. The Angle Orthodontist (2005) 75:1041–1045.[Web of Science][Medline]
Thomas S, Sherriff M, Birnie D. A comparative in vitro study of the frictional characteristics of two types of self-ligating brackets and two types of pre-adjusted edgewise brackets tied with elastomeric ligatures. European Journal of Orthodontics (1998) 20:589–596.
Thorstenson GA, Kusy RP. Resistance to sliding of self-ligating brackets versus conventional stainless twin brackets with second-order angulation in the dry and wet (saliva) states. American Journal of Orthodontics and Dentofacial Orthopedics (2001) 120:361–370.[CrossRef][Web of Science][Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||