The European Journal of Orthodontics 2007 29(1):60-66; doi:10.1093/ejo/cjl063
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Vande Vannet, B.
Right arrow Articles by Wehrbein, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Vande Vannet, B.
Right arrow Articles by Wehrbein, H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2007. Published by Oxford University Press on behalf of the European Orthodontic Society. All rights reserved. For permissions, please email: journals.permissions@oxfordjournals.org.

The use of three-dimensional oral mucosa cell cultures to assess the toxicity of soldered and welded wires

Bart Vande Vannet*, Jean-Louis Hanssens* and Heinrich Wehrbein**

* Department of Orthodontics, Free University of Brussels, Belgium
** Department of Orthodontics, University of Mainz, Germany

Address for correspondence Bart Vande Vannet, Department of Orthodontics, Medical and Pharmaceutical Faculty, Free University of Brussels, Laarbeeklaan 103, 1090 Brussels, Belgium, E-mail: bart.vandevannet{at}pandora.be


    Summary
 Top
 Summary
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusion
 References
 
The aim of the present study was to determine whether there is a difference in toxicity and loss of viability of three-dimensional (3D) reconstructed human oral epithelium (RHOE) cell cultures induced by point-welded (PW), laser-welded (LW), and silver-soldered (SiS) orthodontic wires. Three types of soldered stainless steel (SS) wires: PW, LW, and SiS were prepared (n = 3) and subjected to multiple end-point analysis (MEA). Six pieces were cut from each wire. Each piece was placed on the triplicate cell cultures (RHOE model based on TR 146 cells). After 24 hours of topical exposure, the cell cultures were cut and stained with haematoxylin/eosin. Toxicity was assessed by evaluating the morphological changes and classifying these as mild, moderate, or severe. The in vitro cell cultures were subjected to a 3-(4,5-dimethylthiazol-2-yl)2,5-diphenyl tetrazolium bromide (MTT) assay in order to quantify viability. Copper wires served as the control to determine severe toxicity and native cell cultures were used as a baseline. Untreated SS wire (0.9 mm) was included for comparison with the welded wires.

Histological evaluation with respect to toxicity and measurement of viability in the 3D cell cultures showed no severe toxicity or loss of viability for any of the wires. The morphological ranking of the tested wires from mild to severe toxicity was: SS = PW = LW < SiS. MTT tests revealed the following mean viability values: native cell line (negative control) 98.1 per cent, SS 96.8 per cent, PW 95.5 per cent, LW 95.5 per cent, SiS 85.7 per cent, and copper wires (positive control) 51.2 per cent. Relative comparison between the different welding techniques tested on RHOE revealed, however, that LW and PW wires induced less toxicity/loss of viability compared with SiS wires.


    Introduction
 Top
 Summary
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusion
 References
 
In orthodontics, both soldering and welding are standard methods for the application of auxiliaries and for the modification of force systems by joining wires of different cross-sections (Hannemann et al., 1989Go). The materials used by orthodontists have changed rapidly in recent years and will continue to do so in the future. Disintegration of orthodontic appliances has become a critical issue in orthodontic research because of the potential adverse biological reactions of different materials. Soldered wires, as used in orthodontic appliances, may lead to ionic release and leaching. Oral lesions caused by an orthodontic retainer have been reported (Bishara, 1995Go). Soldered stainless steel (SS) facebows are very susceptible to corrosion (Grimsdottir et al., 1992bGo). Large amounts of copper and zinc and some cadmium from the silver solder have been found. Silver-soldered (SiS) SS wire corrodes more than cobalt–chromium, and releases more nickel and chromium than cobalt–chromium wire (Berge et al., 1982Go). Primarily, as an alternative to soldering, laser welding has been introduced in dental technology (van Benthem and Vahl, 1978Go; Tambasco et al., 1996Go). Incomplete filling of the solder gap, porosities resulting from the production process, poor corrosion properties, and, in particular, a high variability leading to insufficient reliability of soldering techniques, confirm the need for alternative joining techniques, such as laser welding (Hofmann and Lindigkeit, 1999Go; Heidemann et al., 2002Go).

It is speculated that cytotoxic corrosion products from orthodontic appliances might contribute to localized gingivitis (Grimmsdottir et al., 1992a). In order to limit cytotoxic ion release, investigation of the biological performance of orthodontic materials is important (Eliades et al., 2001Go). Other studies indicate that metal ions released from metallic biomaterials may have adverse biological effects at concentrations lower than previously reported (Wataha et al., 2000Go, 2002; Cortizo et al., 2004Go). In animal studies (Gjerdet et al., 1987Go), soldered Co-Cr-Ni wire elicited the most severe reactions, and pronounced cytotoxicity was seen. Leachable toxic components of SiS seem to be of importance in the observed tissue response (Gjerdet et al., 1987Go).

The aim of the present research was to investigate the potential of soldered and welded wires to induce morphological changes in human oral epithelial cell lines and alter the viability as determined by 3-(4,5-dimethylthiazol-2-yl)2,5-diphenyl tetrazolium bromide (MTT) assay.


    Materials and methods
 Top
 Summary
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusion
 References
 

Preparation of the test wires

The orthodontic materials tested in this study were obtained from Forestadent®, Pforzheim, Germany and comprised 0.9 mm SS wire (Forestanit®-wire, spring hard, 0.90 mm 213-6090) and SiS (Silberlot 2.5151 L-Ag72). The composition of the test materials, provided by the manufacturers, is summarized in Table 1. Eight pieces (7 cm in length) were cut, and pairs were united in the middle over a distance of 1 cm by SiS, point welding (PW; Weldman, Bonnel Technology, Schwarzenbruck, Germany) and laser welding (LW; Bego, Dentaurum, Ispringen, Germany). Two pieces of untreated SS wire served as the control. The SiS wire was cleaned and polished, and sterilized before further procedures. From each prepared wire, five 1 mm pieces were cut. This was repeated three times. Copper wire was used as the positive control.


View this table:
[in this window]
[in a new window]

 
Table 1 Chemical composition of the materials tested, as provided by manufacturers.

 

Cell culture

Reconstructed human epidermal and epithelial tissue have been shown to be reliable pre-clinical tools in predicting the irritation potential of topical products (de Brugerolle de Fraissinette et al., 1999; Schaller et al., 2004Go) and these were used in this study.

The reconstituted human oral epithelial (RHOE) tissue models (SkinEthic Laboratories, Nice, France) consist of airlifted, living, multilayered epithelial tissue constructed of oral epithelium. Cultivated in vitro transformed human keratinocytes of the cell line, TR 146 (human cutaneous carcinoma) are produced in polycarbonate inserts in serum-free and chemically defined medium, and form an epithelial tissue (mucosa), devoid of stratum corneum and resembling, histologically, the mucosa of the oral cavity (Rosdy and Clauss, 1990Go). The inserts containing the RHOE tissue models are shipped at room temperature in a multiwell plate filled with an agarose-nutrient solution in which they are embedded. The multiwell plates are sealed with white tape and packed in a sterile aluminium bag. Maintenance medium, cooled with gel refrigerant, is shipped along with the RHOE human tissue models. The cell cultures of oral human tissue and media are standardized for in vitro testing of chemicals or formulations applied on the surface of the tissues or in the medium beneath. Human cells used for the production of the RHOE human tissue models have been tested for biological safety. Absence of HIV integrated proviral DNA, hepatitis C viral DNA, cytomegalovirus DNA, mycoplasma, Hepatitis B antigen HBs, bacteria and fungus are verified by the supplier. The maintenance medium was tested for sterility. However, appropriate safety and sterility precautions were recommended for manipulating human tissue models and media. Each batch of tissues was quality assured by the supplier according to specific quality criteria standards including histology (haematoxylin–eosin staining), tissue viability [MTT mean optical density (OD)] and reproducibility, and tissue thickness (Food and Drug Administration, 1996Go).


Tissue storage

The 24-well plate was opened under a sterile airflow and the sterile filter paper was removed. The cell culture was placed in a 35 mm Ø culture dish (six-well plate) previously filled with 1000 µl of chemically defined maintenance medium (at room temperature). The RHOE was incubated for 24 hours at 37°C (5 per cent CO2) before testing. Prior to testing, the medium beneath each 0.5 cm2 human oral epithelial tissue was changed by adding 500–1000 µl fresh maintenance medium (at room temperature).


Testing procedure

A volume of fresh medium (0.5 ml) was placed under each insert. The different pieces of 1 mm long wire were applied topically onto the surface of the reconstituted epithelial tissues and the culture dishes were placed back in the incubator for 24 hours at 37°C (5 per cent CO2). Untreated cell cultures served as the negative control.


Histological analysis

For histological analysis, the epithelial tissue, with the polycarbonate filter, was cut off the plastic insert with a sharp scalpel and a standard procedure of paraffin inclusion of the fixed cultures (10 per cent formalin) was performed followed by staining with 0.7/1 per cent haematoxylin/eosin. The results are shown in Figure 1. A scoring system was established by taking into account the overall changes in the morphological parameters. Toxicity was assessed by evaluating the morphological changes and classifying them as mild, moderate, or severe compared with the native untreated culture cells (Doucet et al., 1988Go). For control of the reliability of the histological findings, the results were peer reviewed by two independent observers (anatomo-pathologists).


Figure 1
View larger version (72K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 Composite figure of the histological sections.

 

Quantification of cell viability using the MTT assay

The MTT assay is a simple, accurate and reproducible means for measuring the activity of living cells via mitochrondrial dehydrogenase activity. The key component is MTT. Solutions of MTT, in balanced saline, in the absence of phenol red, are ‘yellowish’ in colour. Mitochondrial dehydrogenases of viable cells cleave the tetrazolium ring, yielding blue/purple MTT crystals, which are insoluble in aqueous solutions. Crystals formed by viable cells are retained in polycarbonate filters that are used as substratum for the cultures: an intense purple colour of the cultures shows basal cell viability, whereas the colour remains white when necrosis has occurred. Negative control cultures are dark blue, proof of the viability of a cell, and positive control cultures white/yellow as evidence of cell death (Mosmann, 1983Go; Sjogren et al., 2000Go, Tomakidi et al., 2000). Quantification of cell viability was performed by calorimetric assay. The MTT crystals were extracted by isopropanol, and the OD of extracts was measured at 570 nm (reference filter 690 nm). The results were expressed as a percentage of viability compared with the negative control. Three hundred microlitres of 0.5 mg/ml MTT (Sigma Aldrich, Bornem, Belgium), in defined culture medium, was added to all test wells and incubated for 3 hours at 37°C (5 per cent CO2) on a vibrating platform. After 10 minutes, and 1, and 3 hours, the colour of the wells was observed. After incubation, the MTT solution was discarded. The RHOE was washed twice with 1 ml phosphate-buffered saline pH 7.4 (Life Technologies, Merelbeke, Belgium) at room temperature for 2 minutes, and 1 ml of isopropanol (Merck, Overijse, Belgium) was added to each well. The extraction process was performed over a period of 1 hour at room temperature, on a vibrating platform. Aliquots (200 µl) of blue MTT extraction solution were transferred from each well to the corresponding wells of a 96-well plate. The OD at 570 nm was determined spectrophotometrically using a microplate reader (MIOS, Merck). The mean OD570 of the untreated control tissues exposed to culture medium was set to represent 100 per cent of viability, and the results were expressed as a percentage of these controls. The following formula was used:

Formula


Statistical analysis

Statistical evaluation was undertaken with the Statistical Package for Social Sciences (SPSS®, Chicago, Illinois, USA, version 12.0 for Windows) using the Scheffe's method to test all possible contrasts at the same time. Normal distribution of the data was confirmed and the results were checked for homogenicity.


    Results
 Top
 Summary
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusion
 References
 
In this study, a multiple end-point analysis (MEA), including tissue viability (MTT conversion) and morphological evaluation by histological interpretation, was performed to determine the cytotoxicity of the tested wires.


Histological interpretation

The slides were evaluated and peer reviewed at a magnification x20 and x40 by two histopathologist, who were in agreement (Table 2). The RHOE tissues were sensitive to the majority of the tested wires as there was an indentation at the probe site (Figure 1). The region next to the tested wire was therefore scored. Native cells showed a constant thickness without morphological change. Mild changes were observed for SS, PW, and LW wires. SiS wires showed marked architectural atrophy: most of the upper cell layers were disintegrated but the basal cells remained on the polycarbonate substratum. The ranking of morphological evaluation from mild to severe toxicity was: SS = PW = LW < SiS (Figure 1).


View this table:
[in this window]
[in a new window]

 
Table 2 Peered assessment of the histological interpretation of toxicity.

 

Cell viability with the MTT assay

The qualitative results for the negative control after 10 minutes, and 1 and 3 hours of testing, was blue, indicating that SS, PW, LW, and SiS were non-irritant. The positive control, copper, turned blue/white after 1 hour, evidence of cell death and an acute irritant (Table 3).


View this table:
[in this window]
[in a new window]

 
Table 3 MTT50 values for repeated tests on different cell cultures.

 
The quantitative results were obtained by measuring OD in a calorimeter. To assess the experimental variability, the tested wires were placed in three different wells so that for each wire a mean value could be calculated (Table 4). The mean data (n = 3) of the different wires and controls (n = 6) which reduced MTT–formazan formation by 50 per cent (MTT50) were, respectively, 0.981 (±0.001) for native cells (negative control), 0.968 (±0.01) for SS, 0.955 (±0.01) for PW, 0.955 (±0.01) for LW, 0.857 (±0.01) for SiS, and 0.512 (±0.01) for copper wire (positive control; Table 3). The percentage of viability is shown in Figure 2. SS, PW, and LW demonstrated normal viability. SiS showed a reduction in viability compared with the native cells and the other tested wires. No acute toxicity of the tested wires was found. Comparison of the MTT values showed, as expected, a statistically significant difference between the mean values of the negative control and all wires examined, and the positive control (P < 0.05). SiS resulted in significantly lower viability compared with PW, LW, SS, and native cells, but a significantly higher viability than copper-treated cells (P < 0.05; Table 5). The variation between the repeated tests was very low and explains the significant differences (Table 6).


View this table:
[in this window]
[in a new window]

 
Table 4 Qualitative evaluation of cell viability.

 

Figure 2
View larger version (15K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 Percentage viability of tested wires (obtained by calorimetry).

 

View this table:
[in this window]
[in a new window]

 
Table 5 Descriptive statistical evaluation, dependent variable MTT (post hoc Scheffe test: standard error was 0.008).

 

View this table:
[in this window]
[in a new window]

 
Table 6 Descriptive statistics (dependent variable MTT–Scheffe test).

 

    Discussion
 Top
 Summary
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusion
 References
 
Compared with native cells, SiS wires showed moderate to severe histopathology with, most markedly, a disintegration of the upper cell layers. The thickness of the cell cultures was not preserved. Since MTT is mainly converted by the basal and suprabasal cells in three-dimensional constructs, to detect necrosis of the upper superficial layer of the tissue model, histological tissue analysis is required (De Wever and Charbonnier, 2002Go). This confirmed the necessity of a MEA for evaluation of MTT assays. The collected data give a quantification of the cytotoxicity of the tested wires and can explain the adverse effects seen by other authors (Grimmsdottir et al., 1992a; Hofmann and Lindigkeit, 1999Go; Heidemann et al., 2002Go). Statistical evaluation demonstrated a significantly lower viability for SiS wires compared with the other tested wires. Gjerdet et al. (1987)Go observed moderate to extreme reactions adjacent to soldered joints and around the wire portion, and found that soldered Co-Cr-Ni wire elicited the most severe reactions caused by leachable toxic components of the silver solder. The results of the present investigation are in agreement with the findings of a study designed to assess the cell compatibility of orthodontic wires on a fibroblast culture using the MTT test (Rose et al., 1998Go). Therefore, this RHOE model technique could be used in the future to characterize the effects of dental materials and to reduce the need for animal experiments.

The different histopathology of the SiS and copper wires is in agreement with findings of Wataha et al. (1994)Go who reported that the uptake of Ag+1, Cu+2, and Zn+2 was sufficiently ‘fast’ that in vivo exposure of tissues to these metals for periods of less than 12 hours would be capable of disrupting cellular metabolism.

The biocompatibility of a material is not absolute; it must be measured with regard to the way the material is used. Measuring biocompatibility is a complex process that involves in vitro and in vivo testing. These tests contribute to the understanding of the biological responses to a material but cannot define the biocompatibility of a material with any certainty (Wataha, 2001Go). The RHOE used in this study recreates many of the structural and functional features of the oral mucosa, but is exclusively composed of keratinocytes. For example, Langerhans cells, which play a critical role in cutaneous immune response and cytokine production, are absent from the RHOE (Srivasta et al., 1994Go; Cumberbatch et al., 1996Go). Moreover, the absence of blood-derived and resident leukocytes reduces the complexity of the observable cytokine network. Nevertheless, they are the predominant cell type in human oral mucosa and keratinocytes, and are the first cells in contact with test products, during the initial stages of cellular irritation.


    Conclusion
 Top
 Summary
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusion
 References
 
The RHOE tissue model seems to be suitable to demonstrate the relative toxicity of SiS wires. No acute toxicity, however, could be demonstrated. Whether these initial and early responses of the oral keratinocytes are sufficient to detect and discriminate all types of compounds which are irritant to the oral mucosa remains to be determined. This RHOE model technique combined with a MEA could be used in future experiments to test new materials to reduce the need for animal experimentation. Additionally, once validated, this procedure will be less time-consuming and more cost effective than the usual in vivo procedures. Further research, however, is needed.


    Acknowledgement
 
The authors wish to thank Mrs Marijke Baekeland, Mrs Danielle Blijweert, and Professor Dr E. Wisse [Department of Histology, Free University of Brussels (VUB)], for their assistance and help, and Dr M. Cappadoro and B. De Wever (SkinEthic Laboratories) for their advice and support during this study.


    References
 Top
 Summary
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusion
 References
 

    Berge M, Gjerdet NR, Erichsen ES. (1982) Corrosion of silver soldered orthodontic wires. Acta Odontologica Scandinavica 40:75–79.[Web of Science][Medline]

    Bishara SE. (1995) Oral lesions caused by an orthodontic retainer: a case report. American Journal of Orthodontics and Dentofacial Orthopedics 108:115–117.[CrossRef][Web of Science][Medline]

    Cortizo MC, De Mele MF, Cortizo AM. (2004) Metallic dental material biocompatibility in osteoblastlike cells: correlation with metal ion release. Biological Trace Elements Research 100:151–168.[CrossRef]

    Cumberbatch M, Dearman RJ, Kimber L. (1996) Constitutive and inducible expression of interleukin-6 by Langerhans cells and lymph node dendritic cells. Immunology 87:513–518.[CrossRef][Web of Science][Medline]

    de Brugerolle de Fraissinette A, et al. (1999) Predictivity of an in vitro model for acute and chronic skin irritation (SkinEthic) applied to the testing of topical vehicles. Cell Biology and Toxicology 15:121–135.[Medline]

    De Wever B and Charbonnier V. (2002) Using tissue engineered skin to evaluate the irritation potential of skin care products. Cosmetics and Toiletries Magazine 117:28–36.

    Doucet O, Lanvin M, Zastrow L. (1988) A new in vitro human epithelium model for assessing the eye irritation potential of formulated cosmetic products. In Vitro and Molecular Toxicology 11:273–283.

    Eliades T, Eliades G, Watts DC. (2001) Intraoral aging of the inner headgear component: a potential biocompatibility concern. American Journal of Orthodontic and Dentofacial Orthopedics 119:300–306.[CrossRef]

    Food and Drug Administration. Medical devices; current good manufacturing practice (CGMP) final rule; quality system regulation. Federal Register 61: pp. 52601–52662.

    Gjerdet NR, Kallus T, Hensten-Pettersen A. (1987) Tissue reactions to implanted orthodontic wires in rabbits. Acta Odontologica Scandinavica 45:163–169.[Web of Science][Medline]

    Grimsdottir MR, Hensten-Pettersen A, Kullmann A. (1992a) Cytotoxic effect of orthodontic appliances. European Journal of Orthodontics 14:47–53.[Abstract/Free Full Text]

    Grimsdottir MR, Gjerdet NR, Hensten-Pettersen A. (1992b) Composition and in vitro corrosion of orthodontic appliances. American Journal of Orthodontics and Dentofacial Orthopedics 101:525–532.[Web of Science][Medline]

    Hannemann M, Minarski P, Lugscheider E, Diedrich P. (1989) Materials science studies on the soldering of different orthodontic wires. Fortschritte der Kieferorthopädie 50:506–517.[CrossRef][Medline]

    Heidemann J, Witt E, Feeg M, Werz R, Pieger K. (2002) Orthodontic soldering techniques: aspects of quality assurance in the dental laboratory. Journal of Orofacial Orthopedics 63:325–338.

    Hofmann J and Lindigkeit J. (1999) Festigkeit von lasergeschweißten Drähten für kieferorthopädische Apparaturen. Quintessenz der Zahntechnik 25:437–443.

    Mosmann T. (1983) Rapid colometric assay for cellular growth and survival: application to proliferation and cytotoxicity assays. Journal of Immunological Methods 65:55–63.[CrossRef][Web of Science][Medline]

    Rosdy M and Clauss L-C. (1990) Terminal epidermal differentiation of human keratinocytes grown in chemically defined medium on inert filter substrates at the air-liquid interface. Journal of Investigative Dermatology 95:409–414.[CrossRef][Web of Science][Medline]

    Rose EC, Jonas IE, Kappert HF. (1998) In vitro investigation into the biological assessment of orthodontic wires. Journal of Orofacial Orthopedics 59:253–264.

    Schaller M, Laude J, Bodewalt H, Hamm G, Korting HC. (2004) Toxicity and antimicrobial activity of a hydrocolloid dressing containing silver particles in an ex vivo model of cutaneous infection due to C. Albicans and Methicillin-resistant S. Aureus. Skin Pharmacology and Applied Skin Physiology 17:31–36.[CrossRef]

    Sjogren G, Sletten G, Dahl JE. (2000) Cytotoxicity of dental alloys metals, and ceramics assessed by Millipore filter, agar overlay, and MTT tests. Journal of Prosthetic Dentistry 84:299–236.

    Srivasta BI, Srivasta A, Srivasta MD. (1994) Phenotype, genotype and cytokine production in acute leukaemia involving progenitors of dendritic Langerhans cells. Leukocyte Research 18:499–511.

    Tambasco J, Anthony T, Sandven O. (1996) Laser welding in the dental laboratory: an alternative to soldering. Journal of Dental Technology 13:23–31.

    Tomakidi P, et al. (2000) Assessment of acute cyto- and genotoxicity of corrosion eluates obtained from orthodontic materials using monolayer cultures of immortalized human gingival keratinocytes. Journal of Orofacial Orthopedics 61:2–19.

    van Benthem H and Vahl J. (1978) Tear resistance of laser-welded dental alloys. Deutsche Zahnarztliche Zeitung 33:262–266.

    Wataha JC. (2001) Principles of biocompatibility for dental practitioners. Journal of Prosthetic Dentistry 86:203–209.[CrossRef][Web of Science][Medline]

    Wataha JC, Hanks CT, Craig RG. (1994) In vitro effects of metal ions on cellular metabolism and the correlation between these effects and the uptake of the ions. Journal of Biomedical Material Research 28:427–433.

    Wataha JC, Lockwood PE, Schedle A. (2000) Effect of silver, copper, mercury, and nickel ions on cellular proliferation during extended, low-dose exposures. Journal of Biomedical Material Research 52:360–364.

    Wataha JC, Lockwood PE, Schedle A, Noda M, Bouillaget S. (2002) Ag, Cu, Hg and Ni ions alter the metabolism of human monocytes during extended low-dose exposures. Journal of Oral Rehabilitation 29:133–139.[CrossRef][Web of Science][Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?



This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Vande Vannet, B.
Right arrow Articles by Wehrbein, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Vande Vannet, B.
Right arrow Articles by Wehrbein, H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?